Experimental study of safety in application of the Iris Medical IQ 810 diode laser in clinical treatment of age-related macular degeneration

Zhuravleva, E.S.; Saburina, I.N.; Borzenok, S.A.; Doga, A.V.; Kosheleva, N.V.; Kachalina, G.F.; Magaramov, D.A.; Tonaeva, K.D.

Patologicheskaia Fiziologiia i Eksperimental'naia Terapiia 3: 16-20

2011


ISSN/ISBN: 0031-2991
PMID: 22279734
Document Number: 648061
During the study of diode laser radiation effect in micropulse mode on culture cells of human retinal pigment epithelium it was revealed that the quota of dead cells was a minimum. Besides, a certain conformity between dead cells quota and parameter characteristics of laser radiation. Based on the performed experimental study it was revealed that for a work using the Iris Medical IQ 810 diode laser in the micropulse mode following parameters: duration of pulse set--300ms, duration of function--9.1%, power--750mW are safe for retinal pigment epithelium cells. Rationales of safety in application of the infrared diode laser radiation in micropulse mode in clinic for treatment of age-related macular degeneration (AMD) exampled by cell culture of human retinal pigment epithelium.