Adaptive removal of motion artifact from ECG based on impedance detection

Lin, S.; Lai, L.; Wu, X.

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi 27(3): 529

2010


ISSN/ISBN: 1001-5515
PMID: 20649012
Document Number: 638148
Motion artifact is a significant source of noise in ambulatory physiologic signals and can affect the diagnoses seriously. To cope with this problem, we adopted certain methods for removing motion artifact from ECG signals via synchronous acquisition of the elector/skin impedance as the reference signal of adaptive filter. The results demonstrated that 3rd order RLS adaptive filter can remove the baseline shift effectively with an exciting current of 120 Hz and 50 microA. Such system was implemented to investigate the utility of ambulatory heart rate monitoring, which provides convenient and reliable monitoring services for patients.