Although SYBR Green is used to estimate copy number, its fluorescence varies with amplicon length and adenine/thymine (AT) content. As a result, threshold cycle (Ct) values obtained using real-time polymerase chain reaction (PCR) are lower for longer amplicons (P<0.001) and amplicons with greater AT content (P<0.001). In contrast, neither amplicon length nor AT content affects the Ct with TaqMan probes or LUX-labeled primers.