Experimental study of orbital volume using computer assisted technology

Wang, Y.-X.; Tang, Y.-S.; Shi, J.; Xu, B.

Shanghai Kou Qiang Yi Xue 17(3): 297-299

2008


ISSN/ISBN: 1006-7248
PMID: 18661074
Document Number: 617523
To evaluate the accuracy of the Simplant analytic software for orbital volumetric measurement. The volume of 10 orbits from 5 skulls was evaluated with the Simplant analytic software, and compared with the volume obtained by direct measurement using the water displacement method. The accuracy of volume measurement using Simplant software was assessed statistically by paired t test with SAS6.12 software package. The mean volume was found to be (28.82+/-1.65)ml by direct impression and (28.81+/-1.65)ml by the Simplant analytic software. There was no significant difference between the two methods (P>0.05). The correlation between the techniques was found to be high (r=0.971). The Simplant analytic software for orbital volumetric measurement is a simple and accurate method. Supported by Shanghai Leading Academic Discipline Project (Grant No.Y0203).