Acceptance testing of Computed Radiography systems

Riccardi, L.; Cauzzo, M.C.; Fabbris, R.

La Radiologia Medica 110(5-6): 676-688

2005


ISSN/ISBN: 0033-8362
PMID: 16437053
Document Number: 589731
Acceptance testing of computed radiography systems (CR) is required to verify not only image quality but also compliance with the manufacturer's specifications. Therefore, CR acceptance testing is manufacturer-specific. This paper describes a series of performance tests performed on a large number of CRs (FCR 5000, Fuji) recently acquired by our institution. In particular we describe the following tests: dark noise, uniformity, exposure calibration, linearity and autoranging, limiting resolution, noise and low-contrast resolution, spatial accuracy, laser-beam function, erasure thoroughness, aliasing, grid response. Special attention is given to the practical aspects related to measurement and subsequent image analysis. We report the results obtained in the various tests. No significant variations from the reference levels were found. Nonetheless, in some cases the operating procedures had to be adapted.