Influence of different developing solutions and developing times on radiographic caries diagnosis
Svenson, B.; Petersson, A.
Dento Maxillo Facial Radiology 19(4): 157-160
1990
ISSN/ISBN: 0250-832X PMID: 2097224 DOI: 10.1259/dmfr.19.4.2097224Document Number: 571432
One hundred extracted premolars were radiographed under standardized conditions with D- and E-speed films. The films were developed with conventional and rapid processors using varying developing times. The influence of different developers and developing times on the diagnostic accuracy of radiographic caries diagnosis was evaluated with the aid of ROC technique. Rapid processing using the developing time recommended by the manufacturer led to a lower density and contrast compared with the conventional developer. Underdeveloped films using the rapid processors resulted in an unacceptable low level of diagnostic accuracy, but otherwise it was the same for all developers, developing times and both types of film.