The use of shimstock to assess occlusal contacts: a laboratory study

Harper, K.A.; Setchell, D.J.

International Journal of Prosthodontics 15(4): 347-352

2002


ISSN/ISBN: 0893-2174
PMID: 12170848
Document Number: 550987
This laboratory study investigated the influence of varying occlusal forces, occlusal contacts, or small gaps (0 to 8 microns) on the force required to pull 8-micron shimstock from between a tooth and a restorative material specimen. An apparatus was developed and calibrated to (1) produce and measure very small gaps (0 to 20 microns) and apply variable occlusal forces between specimens, and (2) pull shimstock consistently from between the specimens and measure the removal force. An unworn maxillary molar cusp and polished amalgam were used as the opposing specimens. Combinations of occlusal forces from 0 to 400 g and gaps of 0 (contact), 2, 4, 6, and 8 microns were tested. Shimstock removal forces increased linearly with increasing occlusal force with 0-, 2-, 4-, and 6-micron gap settings. With 0-, 2-, and 4-micron gaps, there was no significant difference in removal force. With a 6-micron gap, the forces rose but at a lower gradient. With the 8-micron gap, the mean force was low (less than 12 g). The shimstock removal force increased with occlusal force and was highest with the smaller gaps. Using 8-micron shimstock, 0-, 2-, and 4-micron gaps cannot be differentiated, a gap of up to 6 microns may still be assessed as a holding contact because the shimstock is still gripped, and even an 8-micron gap may be recorded as a light contact since there is still friction on removal.