Central and juxta-endplate vertebral body screw placement: a biomechanical analysis in a human cadaveric model

Lowe, T.; O'Brien, M.; Smith, D.; Fitzgerald, D.; Vraney, R.; Eule, J.; Alongi, P.

Spine 27(4): 369-373

2002


ISSN/ISBN: 0362-2436
PMID: 11840102
Document Number: 544582
Study Design: In vitro biomechanical testing of transvertebral body screws in different positions in both axial pull-out and toggle. Objectives: To determine the relative strength of unicortical versus bicortical screw fixation within the vertebral body and to determine comparative strength of juxta-endplate and central screw positions with and without staples in both axial pull-out and toggle modes. Summary of Background Data: Loss of fixation is common in centrally placed screws at the rostral end of construct. To preserve segmental vessels, juxta-endplate screw positions are often used. The biomechanical strength of such screw placement methods has not been measured. Methods: Eighty-three human cadaveric vertebral bodies were tested for axial pull-out and toggle with and without staples. Screw positions included central, juxta superior, and inferior endplate. Juxta-endplate screws were toggled in both the rostral and caudal directions perpendicular to the screw axes. Results: Unicortical fixation resulted in a 93% decrease in axial pull-out strength compared with bicortical fixation. Centrally placed screws and juxta-endplate screws were equivalent in axial pull-out if no staples were used. The juxta-endplate screw with a staple that was toggled away from the endplate had the highest yield strength, followed by the central screw with a staple, and then the juxta-endplate screw without a staple toggled away from the endplate. Conclusions: Bicortical fixation is much stronger than unicortical fixation. Centrally placed screws are significantly stronger when used with a staple. When preservation of segmental vessels is desirable, juxta-endplate screws should be placed in such a manner that compressive forces are directed away from the endplate.

Document emailed within 1 workday
Secure & encrypted payments