Dosimetry fundamentals of endovascular therapy using Re-188 for the prevention of restenosis after angioplasty

Kotzerke, J.; Rentschler, M.; Glatting, G.; Schneider, E.; Stabin, M.; Knapp, F.F.; Reske, S.N.

Nuklearmedizin. Nuclear Medicine 37(2): 68-72

1998


ISSN/ISBN: 0029-5566
PMID: 9547753
Document Number: 487836
Various radionuclides can be used for endovascular brachytherapy. A new concept is to inflate the balloon of a dilatation catheter with a radioactive solution. Re-188 can be eluted from a generator system and is available daily. The aim of this study was to obtain dosimetric data for this radionuclide. The dose decrease of Re-188 was calculated and measured with a TLD-system radial to a balloon catheter typically used in cardiology (diameter: 3 mm, length: 20 mm). Using a specific activity of 370 MBq/ml a dose of 0.3 Gy could be achieved within 1 min in a TLD in contact with the balloon. Paired TLDs differed about 3%. A fast dose reduction of 50% and 10% were stated within 0.5 mm and 2.5 mm, respectively. Calculated and measured values were in good agreement. The data are comparable to those known for Y-90. Calculations of dose distribution are consistent with TLD measurements of Re-188. Using a specific activity of 1.85 GBq/ml, a dose of 10-15 Gy at the coronary artery wall can be achieved within 2-3 min. Compared to radioactive stents or wires the use of this liquid beta-emitter is a simple alternative for prevention of the restenosis following the angioplasty.