When is a serum iron really a serum iron? the status of serum iron measurements

Tietz, N.W.; Rinker, A.D.; Morrison, S.R.

Clinical Chemistry 40(4): 546-551

1994


ISSN/ISBN: 0009-9147
PMID: 8149608
Document Number: 437559
Although serum iron determinations play an important role in the diagnostic process, the accuracy of routine methods is suspect. The poor performance of currently used methods is well documented in the quarterly College of American Pathologists survey reports. Compared with the 1990 method of the International Committee for Standardization in Haematology (ICSH), the 1978 ICSH Reference Method shows a +3.4% bias and the standard method of the National Committee for Clinical Laboratory Standards shows a -17.8% bias for iron results. All routine methods checked--DuPont aca, Abbott TDx, Kodak Ektachem (except one lot), BM/Hitachi 717, and Synermed--show a negative bias over the entire analytical range, a significant negative intercept, and extremely poor correlation with the 1990 ICSH method for iron values < 750 micrograms/L. Individual discrepancies of several hundred percent were observed. Imprecision of the methods is not the reason for these discrepancies. We question whether most routine methods measuring low iron concentrations provide results sufficiently reliable for confirmation of iron deficiency.