Biological safety in dentistry: development of a useful method for quality control of sterilization

Arancegui, N.; Lucena, P.H.

Revista Argentina de Microbiologia 26(3): 146-149

1994


ISSN/ISBN: 0325-7541
PMID: 7838979
Document Number: 434496
534 autoclaves from Rosario dentist offices were controlled by a method developed in our laboratory, that consists in: a) a procedures instruction; b) a survey ; c) a colorimetric control; d) a biological control. By this method it is possible to detect the mistake in the autoclave function by only one step. The results showed that 86.90% of the autoclaves lacked thermometers, 76.60% lacked manual thermostats, 83.33% were automatic and 58.80% did not sterilize. It can be concluded the necessity of a periodic control by this method, the importance of a commercial quality control of the furnaces and the urgency of continuous education over biosafety concepts.

Document emailed within 1 workday
Secure & encrypted payments