Diagnosis of metastases in thoracic images. Progress by further development of the conventional film-foil technique?

Lehmann, K.J.; Himmighöfer, U.

Der Radiologe 34(10): 592-598

1994


ISSN/ISBN: 0033-832X
PMID: 7816917
Document Number: 431316
In addition to digital radiography and AMBER, the development of asymmetric screen-film systems is another attempt to optimize chest radiography. Due to reduced contrast in the parenchyma, the former asymmetric screen-film systems did not show sufficient image quality. Three new asymmetric systems with completely different composition are available now. In-Sight VHC (Kodak), High Light GUV (3M) and Opthos D (Agfa) were compared to standard chest films using densitometric curves, a chest phantom for high and low contrast detectability, a nodule detection phantom and patient studies. The sensitivity of nodule detection in the mediastinum has been 41-48% for L-films and 58-65% for the asymmetric screen-film systems. No differences could be demonstrated for nodule detection in the lung field. Contrast in the parenchyma is equivalent to L-films. There is no loss of diagnostic information in the lung field. Differences between the asymmetric systems concern speed, dynamic range and granularity. If AMBER and digital radiography are not available, new asymmetric screen-film systems can improve nodule detection without further investment costs.