Extensive osteolysis around an aseptic, stable, uncemented total knee replacement

Berry, D.J.; Wold, L.E.; Rand, J.A.

Clinical Orthopaedics and Related Research 293: 204-207

1993


ISSN/ISBN: 0009-921X
PMID: 8339482
Document Number: 415359
Extensive osteolysis occurred around an aseptic, well-fixed, stable, uncemented total knee prosthesis. At the time of revision, tibial polyethylene wear and minimal metal-on-metal contact were present. A hypertrophic synovium-like membrane abutted bone in regions of osteolysis. Examination of this membrane revealed polyethylene-wear debris and significant levels of chromium, cobalt, and titanium. Wear debris--polyethylene, metal, or a combination of both--may be responsible for the osteolytic process in this case.