Capillary filtration in venous hypertension. Comparison between the vacuum suction chamber (VSC) device and straingauge phlethysmography
Belcaro, G.; Rulo, A.; Renton, S.
Panminerva Medica 34(4): 151-154
1992
ISSN/ISBN: 0031-0808 PMID: 1293541 Document Number: 392591
To evaluate capillary filtration a group of normal subjects and two groups of subjects with venous incompetence (50 with deep and 50 with superficial incompetence) were studied with the vacuum suction chamber (VSC) device applied onto the internal perimalleolar region. This method was compared with straingauge plethysmography rate of ankle swelling (RAS). By the VSC the time taken for the weal to disappear (VSC time) was considered indicative of capillary filtration. The values obtained with the two methods were well related (r < 0.742) and there was a good separation between patients and normals. The separation between the two groups of patients was significantly better (p < 0.05) with the VSC time. In conclusion VSC time is a good indication of capillary filtration in comparison with straingauge plethysmography RAS. It may be used to assess variations in capillary filtration in venous hypertension and possibly to follow up the effects of treatments.