Selecting a stimulus signal for linear systems analysis of the vestibulo-ocular reflex
Engelken, E.J.; Stevens, K.W.; Enderle, J.D.
Biomedical Sciences Instrumentation 28: 31-36
1992
ISSN/ISBN: 0067-8856 PMID: 1643227 Document Number: 391493
We evaluated 3 types of stimulus signals for use in estimating the transfer function of the vestibulo-ocular reflex. We used individual sine-wave, sum-of-sine, and pseudorandom stimuli. Five normal human subjects were tested 5 times each using each of the 3 stimulus types. Frequency domain techniques were used to estimate the transfer function at 0.01, 0.03, and 0.05 Hz. The most consistent estimates were obtained using individual sine-wave stimuli. The pseudorandom signal yielded the most variable estimates. A sum-of-sine stimulus composed of 3 sine-wave signals provided estimates slightly more variable than the individual sine-wave stimuli, but much less variable than the pseudorandom stimulus. The redundency of the sine and sum-of-sine stimuli seems to be an advantage by providing stable estimates of the transfer function in the presence of noise.