Calibration method for small animal indirect calorimeters

MacKay, S.J.; Loiseau, A.; Poivre, R.; Huot, A.

American Journal of Physiology 261(5 Pt 1): E661-E664

1991


ISSN/ISBN: 0002-9513
PMID: 1951693
Document Number: 382802
A simple mini-butane burner or "artificial rat", for calibration of small animal indirect calorimeters (modified Benedict and Haldane systems), was developed. Substitution of this burner for a live animal at regular intervals in an experiment protocol provides a means by which whole system function can be verified without disturbing respirometer conditions. Heat production, degree of total combustion of butane, proportion of unburned butane, and the CO2 production (.ovrhdot.VCO2)-to-O2 consumption (.ovrhdot.VO2) ratio were determined in animal and burner experiments by direct measurement or were derived from mass equations. Burner security was also discussed. .ovrhdot.VCO2/.ovrhdot.VO2 obtained in calibration experiments using the Benedict system remained constant and was not significantly different from the expected value for total combustion of butane. Observation of a stabilization phase in the Haldane burner experiments illustrated the utility of calibration over the duration of a normal animal experiment. This miniburner gave highly reproducible results and simulated the daily .ovrhdot.VO2, .ovrhdot.VCO2, and heat production of an adult rat.

Document emailed within 1 workday
Secure & encrypted payments