Quantitation of human chorionic gonadotropin with the Abbott IMx immunochemistry analyzer

van Oers, R.J.; Leerkes, B.; Bertschi, I.

Journal of Clinical Chemistry and Clinical Biochemistry 28(1): 59-63

1990


ISSN/ISBN: 0340-076X
PMID: 2179460
Document Number: 365190
The analytical performance of the Abbott IMx automated system was studied and initial results from the human chorionic gonadotropin assay are presented: -Within run and between day precision was good, the majority of all CVs lying between 3 and 5% -Patient results showed excellent comparability with our routine method - No significant specimen-related carry-over was detected - The automated IMx dilution protocol was found to be superior to the manual procedure - The usefulness of the instrument in routine use and as a "Stat" analyzer was very good.