Direct-current potentials created by arcing during monopolar radiofrequency electrosurgery

Tucker, R.D.; Kramolowsky, E.V.; Stasz, P.

Biomedical Instrumentation and Technology 24(3): 212-216

1990


ISSN/ISBN: 0899-8205
PMID: 2344553
Document Number: 356846
Muscle and nerve stimulation during radiofrequency electrosurgery has been a well-known phenomenon for over 50 years. However, as electrosurgery expands into new procedures, such as tissue ablation in the heart, the mechanism of the potentially deleterious stimulation becomes more important. This article presents data showing that large DC currents are generated during electrosurgery and that these potentials increase with increasing sparking at the active electrode. The data also show DC potentials generated by various generators, different operation modes (cut, coagulation, fulguration), and operations on tissues of various impedances.

Document emailed within 1 workday
Secure & encrypted payments

Direct-current potentials created by arcing during monopolar radiofrequency electrosurgery