Practical automation and interpretation of quantitative assays of antibodies to therapeutic proteins, illustrated with human growth hormone

Sportsman, J.R.; Smith, W.C.; Winely, C.L.

Clinical Chemistry 35(8): 1623-1630

1989


ISSN/ISBN: 0009-9147
PMID: 2758631
Document Number: 340276
To quantify concentrations of anti-growth hormone antibody in > 600 serum samples by radioimmunoassay, we devised a system ("BOLAC") to process data and to execute the curve-fitting program LIGAND-PC automatically with an IBM PC-compatible computer. We fit data from each sample to four binding models comprising different combinations of nonspecific binding and one- to two-antibody binding sites. Total antibody concentration is then calculated from the model that is statistically "best." This process occasionally selects a two-binding-site model that severely overestimates the antibody concentration. Errors of this kind are discarded by constraining the product of the second-site antibody's affinity and its concentration to exceed a minimum value (0.05). We evaluated the performance of the BOLAC system by assaying controls and by using computer simulations to demonstrate the high confidence levels attainable in estimation of antibody concentrations. Between-assay variability (CV) was < 25%, and analytical recovery exceeded 90%. These figures are acceptable for an assay based on curve-fitting of competitive radioimmunassay data, allowing clinically relevant assessments of antibody responses in patients' samples. The advantages of the BOLAC system include high throughput and the reporting of results in absolute units of affinities and concentrations.