Enzymic urea assay: a new colorimetric method based on hydrogen peroxide measurement

Lespinas, F.; Dupuy, G.; Revol, F.; Aubry, C.

Clinical Chemistry 35(4): 654-658

1989


ISSN/ISBN: 0009-9147
PMID: 2564817
Document Number: 332513
We describe a new enzymic colorimetric method in which urea is measured in serum by use of a single reagent mixture. Ammonia produced by urea hydrolysis, catalyzed by urease, reacts with glutamate and ATP in the presence of glutamine synthetase. The ADP so produced is assayed in reactions catalyzed sequentially by pyruvate kinase and pyruvate oxidase in a system that generates hydrogen peroxide. The hydrogen peroxide is measured at 500 or 550 nm in a reaction catalyzed by horseradish peroxidase, with phenol/4-aminophenazone as the chromogen. The reaction is complete in 15 min at 37 degrees C. The standard curve is linear up to a urea concentration of 40 mmol/L. Precision is good; CVs ranged from 2.5% to 3.1%. Results by the present method compared well with those by a candidate Reference Method and are not subject to interferences from commonly used drugs and anticoagulants.