Crystaline mineral deposits on copper intrauterine devices. Study involving diffraction by X-rays on whole IUDs

Martin, J.A.; Catau, G.; Chevin, J.C.; Liebaert, F.C.; Schneeberg, J.

Contraception Fertilite Sexualite 16(12): 1049-1056

1988


ISSN/ISBN: 0301-861X
PMID: 12283009
Document Number: 311585
The crystallized mineral deposits on 91 copper IUDs removed by normal procedures were analyzed using x-ray diffraction. 50 Nova-T, 27 Multiload 375, and 14 Gravigard devices were divided into a group of 32 in place for 10-36 months and a group of 59 in place for over 36 months. Studies were performed by x-ray diffraction on whole IUDs in the fresh or dry state and by using classic x-ray, infrared, and atomic absorption spectrometry methods on deposits. Coils of all the IUDs were covered with a layer of cuprous oxide adherent to the metal that flaked over time. Crystals or coatings formed white deposits on 63% of all IUDs, especially when cuprous oxide adhered to the metal. The deposits covered 65-85% of adherent oxides and only 33-38% of flaked oxides for the 1st and 2nd durations of use respectively. Deposits were more abundant on Nova T and Gravigard than on Multiload devices. For the short and long periods of use, the respective frequencies were 67 and 74% on Nova Ts, 25 and 47% on Multiloads, and 88 and 67% on Gravigard. Calcium carbonate (CaCO3) was the only crystalline inorganic compound found, and was present with or without white deposit on 80% of IUDs. It was in the form of calcite, accompanied by vaterite in 40% of cases and rarely by aragonite. Calcium was substituted by magnesium, with an average of 3.7 atoms. Sodium, potassium, and excess magnesium were not in the organic crystalline phase.

Document emailed within 1 workday
Secure & encrypted payments