Determination of beam output factors for the CGR Saturne 2+ linear accelerator beams

Novotńy, J.; Kovár, I.; Jirousek, P.; Hobzová, L.; Wagner, R.

Strahlentherapie und Onkologie Organ der Deutschen Rontgengesellschaft . 164(9): 536-541

1988


ISSN/ISBN: 0179-7158
PMID: 3140402
Document Number: 310402
The demands of a busy clinic require that basic machine calculations be performed as accurately, rapidly and simply as possible. Simple method of predicting X-ray and electron beam output factors for rectangular field sizes on machines that have collimation similar to the CGR Saturne 2+ have been described. The calculated field output factors for rectangular and square field sizes were in agreement with measured data to within +/- 1% for all energies.