The fatigue resistance of orthopaedic wire

Oh, I.; Sander, T.W.; Treharne, R.W.

Clinical Orthopaedics and Related Research 192: 228-236

1985


ISSN/ISBN: 0009-921X
PMID: 3967426
Document Number: 254637
The fatigue resistance of stainless-steel orthopaedic wire was investigated in samples of commercially available (1.2 mm, 16-gauge) stainless-steel wire. The wires were subjected to various types of notching, bending, twisting, and knotting and were then tested on a specially designed wire fatigue testing fixture. The results indicate that a 1% notch is enough to reduce the fatigue resistance of the wire by 63%. Bending, twisting, and knotting of the wire did not appear to seriously affect the fatigue resistance.

Document emailed within 1 workday
Secure & encrypted payments