Intraoperative high-dose-rate 192Ir radical implant in early breast cancer: A quality assurance and dosimetry study
Narayanan, S.Sowmya.; Goel, V.; Sarin, R.; Jalali, R.; Shrivastava, S.Kishore.; Deshpande, D.Dattatray.; Dinshaw, K.Ardeshir.
International Journal of Radiation Oncology Biology Physics 56(3): 690-696
2003
ISSN/ISBN: 0360-3016 PMID: 12788174 DOI: 10.1016/s0360-3016(03)00015-4Document Number: 250721
Purpose: To evaluate the variability in catheter length, geometry, and dosimetric parameters of radical intraoperative high-dose-rate breasts implant during 7-11 days. Methods and Materials: Simulator X-rays, CT scans, and dosimetric studies were repeated on alternate days in 14 consecutive patients treated with radical intraoperative two- or three-plane nylon catheter high-dose-rate implant (34 Gy in 10 fractions within 5 days). Results: Significant variation was found in catheter length, but no major change was noted in implant geometry, homogeneity, or inhomogeneity indexes. A variation in length of >5 mm in one or more catheters was seen in all patients and >10 mm in 11 patients at any time during the implant. Of the 171 catheters in 14 patients, 100 (58%) and 38 (22%) showed a variation of >5 mm or >10 mm, respectively. The variation of >10 mm was reduced from 32% of catheters in the first 5 patients to 17% in the subsequent 9 patients (p = 0.028). Rigid catheter fixation might reduce length variation but may cause skin necrosis if the expanding cavity indents the skin for a long period against the fixation device. A dose homogeneity index of 0.90 (range 0.85-0.92) and dose nonuniformity ratio of 0.20 (range 0.12-0.25) were satisfactory. Conclusion: The catheter fixation and exit catheter length should be measured daily and if the implant is in situ for more than a few days, orthogonal X-rays and, if indicated, dosimetry should be repeated at least once.