An evaluation of the Technicon RA-1000 random-access analyzer

Bromberg, I.L.; Pollard, A.; Cheng, J.; Romaschin, A.

Clinical Chemistry 30(2): 281-283

1984


ISSN/ISBN: 0009-9147
PMID: 6692535
Document Number: 225925
We evaluated the Technicon RA-1000 analyzer, with emphasis on its potential for user-defined method development. Optical linearity and sample pipetting linearity were good. The reagent pipetting system delivered slightly less than the nominal amount, owing to the volumetric effect of added mixing bubbles. Carryover of aqueous solutions was negligible. The instrument had good adaptability for user-defined methods and performed well in method comparisons. The observed dynamic range for enzymes (0-3000 U/L) was excellent. The worklisting software performed its intended functions well, but has limitations. We believe that the RA-1000 represents a significant contribution to the practice of clinical chemistry. A sophisticated benchtop machine, it includes several innovations, along with a few problems that are peculiar to its technology.