Dialysis leucopenia--no correction after prolonged washing of the membrane

Hoenich, N.A.; Laing, C.; Buckley, P.; Woffindin, C.; Kerr, D.N.

International Journal of Artificial Organs 6(3): 113-114

1983


ISSN/ISBN: 0391-3988
PMID: 6874122
Document Number: 215665
Re-use of hemodialyzers diminishes the extent of leukopenia observed during hemodialysis. A possible cause of this change is the prolonged washing of the membrane by an aqueous solution on both sides and by protein contained solutions on the blood side which removes some substance from the membrane which affects leukocytes directly or by activation of complement. A series of studies to test this hypothesis were performed using 2 hemodialysis membranes.sbd.Cuprophan (Enka AG) and Cellulate (Cordis Dow Corporation).sbd.and the results obtained demonstrate no elution of any H2O soluble substance from the dialyzer membrane.