Immunochemiluminometric assay for hepatitis B surface antigen
Schroeder, H.R.; Hines, C.M.; Osborn, D.D.; Moore, R.P.; Hurtle, R.L.; Wogoman, F.F.; Rogers, R.W.; Vogelhut, P.O.
Clinical Chemistry 27(8): 1378-1384
1981
ISSN/ISBN: 0009-9147 PMID: 7273398 Document Number: 178106
A novel "sandwich" immunoassay for hepatitis B surface antigen monitored by chemiluminescence is described. The method involves use of an antibody-coated microtitration plate and requires 100 micro L of test specimen. The antigen binds to the antibody during the first 2-h incubation and, after an intermediate wash step, the sandwich is completed by 2-h incubation with antibody to antigen that has been labeled with an isoluminol derivative. A final wash step follows. A luminometer, built in-house, adds "microperoxidase" and peroxide, to initiate chemiluminescence, and provides automated readout at 10-s intervals. Results compare well in specificity and sensitivity with those of a comparison radioimmunoassay procedure. Within- and between-assay variability (CV) is 7 to 13% (n = 6). All reagents are stable at 4 degrees C for at least several months. Use of a non-radioisotopic label in this assay avoids the stability problems and inconvenience associated with radioactivity.