Comparative study between the classical Westergren and the new sealed vacuum extraction methods for erythrocyte sedimentation rate determination

Wiwanitkit, V.; Siritantikorn, A.

Journal of the Medical Association of Thailand 84(4): 577-580

2001


ISSN/ISBN: 0125-2208
PMID: 11460972
Document Number: 1468
A comparative study between the classical Westergren and the new sealed vacuum extraction methods for erythrocyte sedimentation rate determination was performed on 80 individual subjects. The comparison-of-methods plot sealed vacuum extraction method (X) vs Westergren method (Y) gave a least square linear regression equation of Y = 1.001 X + 0.03 (r = 0.99) and precision analysis gave a coefficient of variation below 3 per cent. Compared to the classical method, the sealed vacuum extraction method was easy to use, showed acceptable precision and reduced biohazard risk to the practitioner, therefore, it was suitable for a laboratory setting with a high rate of blood-borne infectious diseases such as Thailand.