Background: The recent identification of murine leukemia virus (MLV)-related viruses in patients with chronic fatigue syndrome (CFS) has aroused much interest, not least among sufferers. However, other studies failed to detect these viruses in CFS patients. Methods: We wanted to establish a MLV-related virus real-time PCR for routine diagnostics. Results: Our study identified false positive MLV-related virus results due to a contamination of Superscript III Platinum One-Step Quantitative RT-PCR System (Invitrogen). Conclusions: This observation may be helpful to elucidate discrepant results for the detection of MLV-related virus like xenotropic MLV-related virus (XMRV) in recently published studies.