Exit block: a postoperative complication of pacemaker implantation

Laczkovics, A.; Stellwag, F.; Domanig, E.

Wiener Klinische Wochenschrift 90(17): 633-636

1978


ISSN/ISBN: 0043-5325
PMID: 695654
Document Number: 135455
Exit block occurs in 7% of cases following implantation of a pacemaker for the first time and appears to be a serious and unpredictable occurrence which cannot be prevented by the selection of a particular technique or type of electrode of pacemaker. Therapeutic measures in patients who are not dependent on a pacemaker consist primarily of exchanging the intracardial lead. In all patients, as well as in those cases with recurrence of exit block, a pulse width-adjustable pulse generator should be implanted without resorting to other methods of questionable efficacy. The implantation of such a pulse generator was successful in all our cases.