Evaluation of a New Procalcitonin Assay for the Siemens ADVIA Centaur® with the Established Method on the BRAHMS Kryptor®

Robert-Sanders, J.; Schoorl, M.; Dekker, E.; Snijders, D.; Boersma, W.G.; Boekel, E.T.

Clinical Laboratory 57(5-6): 415-420

2011


ISSN/ISBN: 1433-6510
PMID: 21755834
Document Number: 12495
Background: In this study, we compared the BRAHMS Kryptor procalcitonin (PCT) assay with the newly developed ADVIA Centaur BRAHMS PCT assay. Furthermore, the long-term stability of PCT at - 20 ºC was assessed. Methods: Samples from 97 patients with lower respiratory tract infections were retested on both systems and compared with Passing-Bablok regression over two clinically relevant cutoff ranges for PCT, 0 – 2.0 μg / L and > 2.0 μg / L. Results: After storage for 2.5 to 4 years, PCT levels in patient sera declined only 3.7 %. Passing-Bablok regression analysis of the total sample range (n = 97) showed that both methods correlated well (r = 0.9944), although with a deviation from the line of identity (y = 0.880x – 0.025 μg / L). Comparison of both methods within the clinically important interval of 0 – 2.0 μg / L showed acceptable correlation (y = 0.943x + 0.010 μg / L). Conclusions: The ADVIA Centaur BRAHMS PCT assay showed good correlation with the established Kryptor method. Therefore, this new technique can be used in clinical routine with the same clinical interpretation.

Document emailed within 1 workday
Secure & encrypted payments

Evaluation of a New Procalcitonin Assay for the Siemens ADVIA Centaur® with the Established Method on the BRAHMS Kryptor®