Evaluation of a New Procalcitonin Assay for the Siemens ADVIA Centaur® with the Established Method on the BRAHMS Kryptor®
Robert-Sanders, J.; Schoorl, M.; Dekker, E.; Snijders, D.; Boersma, W.G.; Boekel, E.T.
Clinical Laboratory 57(5-6): 415-420
2011
ISSN/ISBN: 1433-6510 PMID: 21755834 Document Number: 12495
Background: In this study, we compared the BRAHMS Kryptor procalcitonin (PCT) assay with the newly developed ADVIA Centaur BRAHMS PCT assay. Furthermore, the long-term stability of PCT at - 20 ºC was assessed. Methods: Samples from 97 patients with lower respiratory tract infections were retested on both systems and compared with Passing-Bablok regression over two clinically relevant cutoff ranges for PCT, 0 – 2.0 μg / L and > 2.0 μg / L. Results: After storage for 2.5 to 4 years, PCT levels in patient sera declined only 3.7 %. Passing-Bablok regression analysis of the total sample range (n = 97) showed that both methods correlated well (r = 0.9944), although with a deviation from the line of identity (y = 0.880x – 0.025 μg / L). Comparison of both methods within the clinically important interval of 0 – 2.0 μg / L showed acceptable correlation (y = 0.943x + 0.010 μg / L). Conclusions: The ADVIA Centaur BRAHMS PCT assay showed good correlation with the established Kryptor method. Therefore, this new technique can be used in clinical routine with the same clinical interpretation.